Ring comparisons

Comparative measurement in the relative humidity range: Hygrometer

A comparative measurement is carried out for the measured variable relative humidity. A humidity sensor with measuring device is sent as the calibration item. The calibration item must be calibrated in accordance with guideline DKD-R 5-8:2019 sequence B. The deviation of the relative humidity at the following measuring points must be determined:

  • 10 % rH, 50 % rH, 90 % rH each at 10 °C
  • 10 % rH, 50 % rH, 90 % rH each at 23 °C
  • 10 % rH, 50 % rH, 90 % rH each at 70 °C

The assigned reference values are determined using the weighted average of all participant results from the accredited calibration laboratories. The results are evaluated using the En‑value for the assigned value.

Schedule and procedure:
Expression of interest: by 21 June 2024
Planned start ring comparison: August 2024

The comparison measurement takes place in a ring, whereby a calibration is carried out in the pilot laboratory at least before and after the calibration at the participants. Each participating laboratory has two calendar weeks to carry out the calibration and is responsible for the insured forwarding of the calibration items. An exact schedule will be sent to the participating laboratories before the start.

Are you interested in our next ring comparison? Then fill out the contact form below by 21 June 2024. We will then send you a non-binding offer. Participation is considered binding as soon as the offer is accepted and an order confirmation is sent.

Note: A minimum number of eight participating laboratories must be reached in order to carry out this interlaboratory comparison.

Captcha image

A summary of the comparison measurement can be found in the description.